- Dual channel (4-lead) all digital combo TENS, EMS, and Microcurrent (MENS)
- Five custom, preset user programs
- Each program offers 10 treatment phases
- 4 TENS modes (Burst, Normal, Modulation, Dual Frequencies)
- In TENS mode, Channels 1 and 2 can be programmed independently
- Last setting recall and parameter lock
- Electrode detection function
- Optional AC wall adapter
- LCD display shows all parameters
- Treatment timer (1-30 minutes)
- Low battery power detection
- Electronic self diagnosis test for safety
- 2 year warranty
- Comes with 1 pkg of 4 reusable self-adhering electrodes, 1 set 43” pin lead wires, 1 each 9V heavy duty battery, vinyl protective case, and instruction manual. AC adapter sold separately.
- Leadwire reorder # LW2152, #LW2153
- Channel: Dual, independently programmed in TENS mode
- Output Current: Constant current, peak value display
- Amplitude TENS/EMS: Max 80 mA at 500 ohm load
- Microcurrent: Max 750 µA at 50K ohm load at maximum output setting
Max. charge per pulse / Max. average current
TNS - 20 µCoulomb / 4 mA
EMS - 24 µCoulomb / 2.4 mA
MCR - 187 µCoulomb / 0.3 mA
- Pulse Width:
TENS 50 ~ 250 microsec
EMS 100 ~ 300 microsec
Microcurrent 1 ~ 250 microsec
- Pulse Rate:
TENS 1 ~ 200 Hz; pulse rate and width can be set independently on each channel on constant mode
EMS 25 ~ 100 Hz
Microcurrent 0.3 ~ 400 Hz
- Modes:
1. TENS - Constant, Burst, Modulation, Dual Frequency
2. EMS - Hold time: 0.5 ~ 30 sec, Rest time: 1 ~ 60 sec.; Ramp up/Ramp down time 0.1 ~ 6 sec.
3. Microcurrent - Constant only
4. Program - up to 5 programs
- Programming Capability: Up to 5 programs, Steps per program: 10
- Timer: 30 minute digital timer, with accuracy of +/- 2%
- Display: LCD (46 mm x 37 mm)
Other Functions Memory of the previous setting parameters and mode, keyboard lock, automatic power off, low battery power detection, abnormal pulse output detection and electronics self-test
- Power Source: Dual power source with DC 9V battery or AC adapter
- Classification: Class II/ Internally powered equipment, TYPE BF
Conformity CE Marking
- Size: 11.5 cm x 6.9 cm x 2.7 cm
- Weight:185 grams including battery
|
|